Keywords: Fuel cell membrane, mechanical property, modulus, coefficient of hygroscopic expansion, humidity control
ABSTRACT
Fuel cells use thin polymer films as the solid electrolyte. The polymers used in this application must have desired properties such as high proton conductivity,high chemical and thermal stability, and low gas permeability. It is important to have a good understanding of the physical properties of electrolyte membranes, especially under different operating temperature and humidity conditions.
An enhanced relative humidity accessory on TA Instruments Discovery Hybrid Rheometer is introduced for the analysis of fuel cell membrane. The measurement results provide quantitative evaluation to the mechanical properties (modulus, tan delta) of the membrane as a function of both temperature and humidity. In the meantime, the rheometer is also capable of measuring the coefficient of hygroscopic expansion (CHE) of the membrane.
INTRODUCTION
PFSA membranes are based on the Short Side Chain (SSC) perfluorosulfonic acid (PFSA) ionomer and have excellent lifetime for fuel cells,electrolyzers, and various electrochemical technologies. PFSA membranes show good proton conductivity, high gap barrier, and are capable of maintaining physical properties within a wide range of operation temperature and humidity. Water management is crucial in fuel cell applications. The PFSA copolymer is reported to have strong water retention within the membrane and performs well in low-humidity electrochemical device operation.
A dynamic mechanical analyzer (DMA) with humidity control capability is an ideal instrument for quantitative evaluation of the physical properties of fuel cell membranes. Figure 1 shows TA Instruments DMA850 with the humidity control accessory attached. This humidity chamber is capable of controlling the relativity humidity from 5 to 95% with a wide range of operating temperatures from 5 °C to 120 °C.
Figure 1. TA Instruments DMA850 with the relative humidity accessory
In addition to using a traditional DMA instrument, TA Instruments rotational rheometers (i.e., the Discovery series Hybrid Rheometers) can also perform dynamic mechanical analysis on solid films. In this application note, the TA Discovery Hybrid Rheometer with a Relative Humidity (RH) accessory (shown in figure 2) is used for the mechanical property measurements of a PFSA membrane. This rheometer uses the same humidity generator as that is used on the DMA 850. Its operational specifications are listed in figure 3. The modulus and dampening factor of the PFSA membrane are measured under both temperature and humidity control.
Figure 2. TA Instruments Discovery Hybrid Rheometer with the relative humidity accessory
Figure 3. The relative humidity accessory operational specifications
EXPERIMENTAL
Commercial PFSA film was used in this research. The film is transparent with thickness of about 50μm. A TA Instruments Discovery Hybrid Rheometer (HR-30) equipped with the relative humidity chamber was used for the measurements. The PFSA film sample was loaded onto a film tension test geometry as shown in figure 4.
Figure 4. The film/fiber tension geometry on Discovery Hybrid Rheometer
Dynamic mechanical tests were performed in 2 different scenarios with both temperature and humidity control. In the first set of tests, relative humidity was held constant at 30%. The temperature was increased stepwise from 10 °C up to 90 °C. The modulus and dimension change of PFSA film with increasing temperature were monitored as a function of temperature. In the second set of tests, temperature was held constant at 30 °C, 50 °C and at 90 °C. The relative humidity was increased stepwise from 5% up to 90%. In both scenarios, dynamic oscillatory tests were conducted at a frequency of 1 rad/s and with a strain amplitude that is within thelinear viscoelastic region of the sample.
RESULTS AND DISCUSSIONS
Figure 5 shows the results of an iso-hume temperature step test. The relative humidity was held constant at 30%. Dynamic mechanical measurement was performed at a frequency of 1 rad/s with an oscillation amplitude of 0.1%, which is within the linear viscoelastic region of the sample. The measurement results show that sample modulus (E’) is decreasing with increasing temperature. In the meantime, it is observed that the length (i.e. gap) of the film is increasing with temperature due to thermal expansion.
Figure 5. Dynamic mechanical measurement performed under a constant relative humidity of 30%. The temperature is increased stepwise from 10 °C to 90 °C. The recorded geometry gap signal is the length of the film sample
The mechanical properties of PFSA film can also be measured at an isothermal temperature but with a change in relative humidity (shown in figure 6). In this measurement, the test temperature was held constant at 30 °C, and the relativity humidity was increased stepwise from 10% up to 90%. Dynamic testing was conducted at 1 rad/s frequency with an oscillation amplitude set at 0.1%. The measurement results show that the modulus of PFSA film is decreasing with increasing environmental humidity. When water molecules diffuse into the film, it acts as a plasticizer to reduce its stiffness. Therefore, higher humidity results in a lower storage modulus (E’). Also, the length of the film is observed to increase with increasing humidity.
Figure 6. Dynamic mechanical measurement performed under an isothermal temperature of 30 °C. The environmental humidity is increased stepwise from 10% up to 90%. The recorded geometry gap signal is the length of the film sample
From both the isothermal and iso-hume test results, one can see that this film takes a certain amount of time (usually longer than 1 hour) to reach equilibration at each temperature and humidity step. Figure 7 shows the summary of sample equilibration modulus (E’) change as a function of humidity. The results show that under both test temperatures of 50 °C and 90 °C, E’ decreases with increasing environmental humidity. Also, higher temperature results in lower E’ at the same humidity level.
Figure 7. Dynamic mechanical tests of PFSA film. Monitoring the changes of storage modulus (E’) as a function of environmental humidity. Blue: 50 °C and Red: 90°C.
During the dynamic measurements, the changes of sample dimension (i.e. length) can also be monitored along with the measurement of modulus. Figure 8 shows the measurement results of sample length change (i.e. delta L) as a function of changing humidity at 50 °C and 90 °C. This provides the capability of quantifying the coefficient of hygroscopic expansion (CHE). In general, CHE is described as the change of length per meter per
%RH, which correlates to the slop of the delta L over %RH. The results shown in figure 8 indicate that PFSA film has a different CHE at different measurement temperatures.
Figure 8. Dynamic mechanical tests of PFSA film. Measuring the coefficient of hygroscopic expansion at different temperatures. Blue: 50 °C and Red: 90 °C.
Another method for CHE measurement is the iso-force humidity step. In this iso-force humidity step test method, a constant low static force (e.g. 0.1 N) is applied to the sample while equilibrating or changing humidity. The benefit of using this iso-force method is that the sample is under a much lower tension force compared to dynamic testing. This minimizes the potential of the sample creeping due to the additional applied oscillation force, so the sample length change can be measured more precisely.
REFERENCES
1. A Baroutaji; JG Carton; M Sajjia; AG Olabi. “Materials in PEM Fuel Cells”. 2016 Elsevier Inc. DOI: 10.1016/B978-0-12-803581-8.04006-6
ACKNOWLEDGEMENT
This paper was written by Dr. Tianhong (Terri) Chen, Principal Applications Support Scientist at TA Instruments.
更多精彩内容您可以通过
TA仪器小程序--文档
找到您想要的文章
07-01 英斯特朗
连载 | 药物一致性评价与粒度分析(三)07-01 欧美克仪器
【仪器百科】LS-909丨干湿二合一激光粒度分析仪07-01 欧美克仪器
标准物质解决方案 | PFASs(全氟及多氟化合物)06-29
第九期阿尔塔有约 | 环境专题【新污染物:PFAS】技术研讨会精彩回顾及提问解答06-29
“绿色技术范式”,分析化学未来发展方向——访中国分析测试协会副理事长、辽宁省分析科学研究院原院长刘成雁教授06-29 转载仪器信息网
华西医院-标准型数显脑立体定位仪、双通道体温维持仪、体式显微镜安装完成06-29 迈越生物
科鉴检测助力2家仪器企业获得首批产品可靠性认证证书06-28 科鉴检测
德国耶拿:锂电池生命周期分析解决方案06-28 德国耶拿
AI已来!生命科学本科教学如何紧跟技术浪潮06-28 Opentrons
盛瀚售后,五星级服务的秘诀是什么?06-28 SHINE
专为汽车制造商打造的柔性解决方案——实现制程控制06-28
西北工业大学-脑立体定位仪安装完成06-28 迈越生物
会议邀请 | 第九届海上检验医师论坛06-28
卓立要闻 | 创新发展ing…6月卓立“大事小情”速览06-28 光电行业都会关注
打造信任合作伙伴!2024年度卓立汉光客户满意度调查开启06-28 光电行业都会关注
如何挑选适用于三阶光学非线性的测量系统?Z扫描测量系统来助力!06-28 光电行业都会关注
招聘启事—中国科学院沈阳自动化研究所微纳光学测量表征技术课题组06-28 光电行业都会关注
谱育科技作为主要完成方 荣获2023年度国家科学技术进步一等奖和二等奖06-28 点击关注→
仪器原理丨顶空仪与吹扫捕集仪科普小知识06-28 天美色谱