* Includes non-linearity hysteresis and non-repeatability at a fixed temperature ** An input of 18-24VDC/AC is required to drive a 500Ω load; the same input is also required for a 1-10VDC output *** 3-wire 4-20mA
Accuracy*:
<+/-1.0% FSO
Stability:
+/-0.5% FSO/yr
Thermal Effects:
+/-0.075%/°C
Overpressure:
1-10”WC = 100”WC
11-20”WC = 10x rated pressure
21-30”WC = 5x rated pressure
Over 30”WC = 45 PSI
FSO Pressure Range:
0/0.25”WC up to 0/15PSI
(bi-directional ranges available)
Compensated Range:
10° to 50°C
Media:
Non-corrosive, non-ionic gases
Operating Humidity:
90% RH non-condensing
Operating Temp:
-25° to 70°C
Input Supply**:
12-24 VDC/24VAC
Supply Current:
<30mA
Load Resistance:
2KΩ minimum on voltage out, 250Ωmaximum loop resistance on 4-20mA output
(500Ω max loop resistance available upon request**)
Output Signal:
1-5VDC, 1-6VDC, 1-10VDC** or 4-20mA*** (call for custom outputs)
Circuit Protection:
Output limited and reverse voltage protected
Electrical Connections:
Pluggable Screw Terminal Block
Pressure Connections:
Barbed fitting for 1/8” ID tubing
Enclosure:
Impact resistant ABS plastic
Dimensions:
Approx. 3.5” x 2.7” x 1.4” (8.7cm x 6.8cm x 3.5cm)