Lowest-cost calibration to -196 °C
If you need to do calibrations at the triple point of argon but don’t
want the complexity and cost of using an argon triple point cell, Fluke
Calibration's Model 7196B LN2 Comparison Calibrators will
solve your problems. And they do it for less than half the price of
other argon triple point simulators.
The nominal boiling point
of nitrogen is –196°C at one atmosphere of pressure. The defining triple
point of argon is –189.3442°C. While there is a difference between the
nominal boiling point of nitrogen and the argon triple point, the
difference can be corrected for mathematically, and an uncertainty of
less than 2 mK from the actual argon triple point is achievable.
Fluke Calibration's LN2 Comparison Calibrators consist of a
vacuum-sealed stainless steel Dewar Flask, a high-purity copper block,
and a precision-fit lid. The dewar is filled with LN2 and the
copper block is suspended in it; an SPRT is inserted into the block and
a calibration is performed against your own calibrated SPRT. The 7196B-4 includes four 8-mm (0.32") wells. The 7196B-13 includes five 8-mm (0.32") wells and eight 6.35-mm (0.25") wells.
Fluke Calibration's LN2 Comparison Calibrators are neither
expensive nor complicated to use. If you need supporting data or would
like to discuss the theory of operation of an LN2 Comparison Calibrator, call Fluke Calibration today. (Or come to one of our calibration training courses and we’ll show you.)
Specifications |
Temperature | Nominal –196°C depending on atmospheric pressure |
Thermal Wells | 7196B-4: four 8 mm (0.32") I.D. wells
7196B-13: five 8 mm (0.32") I.D. wells, eight 6.35 mm (0.25") I.D. wells
Both blocks: 275 mm immersion from top of lid to bottom of well, 150 mm immersion into copper block |
Dimensions (outer diameter x height) | 168 mm x 406 mm (6.6 in O.D. x 16 in) |
Temperature Stability | Typically better than 2 mK/20 min[1] |
Temperature Uniformity | <0.0004 °C |
Volume | 5.47 liters (1.45 gallons) |
[1] The equilibrium temperature will change with the
atmospheric pressure (dT/dp » 0.085 mK/Pa). The actual temperature
stability depends on atmospheric pressure stability. |