Standard models and available pump laser power matrix
TiF-100ST (standard wavelength range)
TiF-100LW (extended IR range)
Pump laser power
Wavelength tuning
Output power
Wavelength tuning
Output power
3 W
740-860 nm
>350 mW аt 800 nm
n/a
n/a
4 W
735-870 nm
>500 mW аt 800 nm
n/a
n/a
5 W
730-875 nm
>650 mW аt 800 nm
830-950 nm
>200 mW at 950 nm
6 W
725-880 nm
>850 mW аt 800 nm
830-960 nm
>200 mW at 960 nm
8 W
720-885 nm
>1.25 W аt 800 nm
830-970 nm
>200 mW at 970 nm
10 W
715-890 nm
>1.55 W аt 800 nm
830-980 nm
>150 mW at 980 nm
It is possible to combine both ranges in a single laser system, please enquire for details.
General optical specifications
Output pulse duration1)
<100 fs
Spectrum width1) (FWHM)
>8 nm
Pulse repetition rate (fixed)
80±10 MHz
Pump laser
integrated or stand-alone DPSS 532 nm low-noise laser (up to 10 W)
Beam mode
TEMoo (M^2 <1.2)
Beam diameter (at 1/e^2)
<2 mm
Output polarization
linear, horizontal, PER >20 dB
Beam divergence
<1 mrad
Long-term stability2)
<0.1% rms
Noise3)
<0.2% rms
Physical dimensions (L × W × H)
Laser head
516 × 350 × 118 mm (including integrated pump laser)
Pump laser control unit
353 × 360 × 119 mm
Closed-loop chiller
430 × 340 × 190 mm
Environmental and utility specifications
Operating temperature
15-30 °C
Relative humidity
<60%, non-condensing
Voltage
single-phase; 100-240 VAC; 50/60 Hz
Power consumption
<2 kW
Available configuration packages4)
Thermally stabilized monolithic body
included in any package
Pulse train SMA output with mode-lock LED
included in any package
“Auto” factory package
- built-in spectrometer - single-touch wavelength tuning w. presets - built-in power meter - active output power stability locking - Windows software with configurable widgets - automatic mode-lock start and monitoring PC requirements: USB 2.0 port, Windows 10
"Basic" factory package (default)
- push-button starter - USB 2.0 wavelength tuning via step-motor slit (via step number information and calibration) PC requirements: USB 2.0 port, Windows 10
1) - when tuned to 800 nm central wavelength; 2) - after 30 min warm-up with cold start, during 12-hour continuous operation under equal room temperature conditions using recommended stabilized closed-loop chiller with proper capacity and recommended low-noise on-board integrated highly stable pump laser with active power locking; 3) - measured from 10 Hz to 10 MHz; 4) - please select one of the packages for your system; certain features may be tailored or combined differently according to specific customer requirements.