The STEM Integrity 10 Reaction Station enables you to conduct 10 different reactions simultaneously within the same reaction unit, each reaction being conducted within its own cell, at its own individual temperature and stir rate.
The STEM Integrity 10 Reaction Station is an excellent screening tool that can be used in most laboratories, with:
• 10 individual cells in one reaction block • Individual control of temperature and stirring rate for each cell • Temperature range of -30°C to 150°C • Stirring rate of 350 rpm- 1200 rpm • Cell working volume of 2ml- 25ml • Optional attachments for refluxing, and working under vacuum or inert gas conditions • Optional multi-infrared probes for solubility/crystallisation studies • Automatic microprocessor control through a touchscreen
If desired, fast heating and cooling rates can be selected, with temperature ramps of between 0.1°C/min to 5°C/min. There is also a crash function for even faster temperature changes, which is ideal for kinetic determinations.
Extreme temperature accuracy is guaranteed, with a temperature stability of ± 0.2°C and a maximum set point overshoot of 0.1°C. The temperature may be maintained either through the block itself or by using a probe within each cell's solution.
This accuracy can be maintained over a wide temperature range of -30°C to 150°C, with precise, independently controlled temperature profiles, and homogeneous sample mixing may be assured with stirring rates of between 350rpm to 1200rpm using magnetic stirrers.
With working volumes of between 2ml to 25ml, the STEM Integrity 10 is an excellent screening tool for most laboratories and can also be used to establish ideal process conditions.
Solubility tests using turbity measurements
The STEM Integrity 10 Reaction Station is available in 2 formats depending upon your needs:
• Option 1: STEM Integrity 10 Reaction Station by itself
• Option 2: Clarity System, which consists of the STEM Integrity 10 Reaction Station, with additional solubility and crystallisation evaluation apparatus and software.
Intrusive or non-intrusive IR turbidity probes are available for the accurate determination of transitions, providing further functionality to the solubility process.
Technical Specification
Number of positions
10
Cell cavity diameter
25.5 mm
Glass vessel fill level
2 - 25 ml
Temperature range
- 30°C to + 150°C
Temperature difference between any two positions
180°C
Temperature overshoot
0.1°C (maximum)
Maximum controlled heating rate
5°Cmin
Maximum controlled cooling rate
5°Cmin
Stir speed range
350 - 1200 rpm
Viscosity capacity
Glycerine at 25°C
Recommended stir bars
12 4.5 mm (cylindrical) or 10 6 mm (oval)
Measurement
Temperature Range
- 40°C to + 160°C
Temperature Resolution
0.01°C
Temperature Accuracy
± 0.5°C
Stirrer Speed range
350 - 1200 rpm
Stirrer Resolution
1 rpm/min
Stirrer Accuracy
± 10 rpm/min
Measured external temperature (optional thermometer) range
Range
- 40 °C to + 160°C
Resolution
0.01°C
Accuracy
± 0..5°C
Ordering Information
Part Number
Description
PS200000
Integrity 10 with 10 individually controlled cells, 230V
PS200000X1
Integrity 10 with 10 individually controlled cells, 110V
PS200000X6
Integrity 10 with 10 individually controlled cells, 230V, EU Plug
ATS20001
Integrity 10 reflux unit with inerting caps
ATS10075
Glass tubes 24 150 mm, 22 thread pack of 10
AZS4206
Stirrer bars 10 6 mm, pack of 10
AZS4235
Stirrer bars 12 4, 5 mm, pack of 10
ATS10001
Multi-Temp 10 module
ATS10027
Thermocouple probe, pack of 6
ATS11005
Integrity software
ATS10360 1
Non Intrusive IR sensor, pack of 1
ATS10360 5
Non Intrusive IR sensor, pack of 5
ATS10360 10
Non Intrusive IR sensor, pack of 10
ATS10230
Intrusive IR probe in stainless steel PID-NIR5-BNSD, pack of 1
ATS10230H
IR probe in Hastelloy (nickel-based alloy with high corrosion resistance)