DSC扫描过程中缓冲液的选择

发布时间: 2016-02-29 11:36 来源: TA仪器
领域: 化工试剂/助剂
样品:缓冲液项目:DSC扫描
参考:Christin T. Choma TA Instruments

方案文件名 下载

DSC扫描过程中缓冲液的选择

下载此篇方案

Although it is often assumed that the pH of a buffer solution remains constant during a differential scanning calorimetry (DSC) experiment, the pK of a buffer can in fact change signifi cantly between 0 oC and 100 oC. Such pH changes can result in sample denaturation, aggregation and precipitation that, in the absence of knowledge about the temperature stability of the buffer, could erroneously be ascribed solely to temperature-induced unfolding of the biomolecule.