用四极杆飞行时间质谱实现高通量鸟枪法脂质组学

发布时间: 2014-03-11 16:45 来源:华质泰科生物技术(北京)有限公司
资料类型:PDF
资料文件名 下载

Ejsing_Ekroos_JChromB_Shotgun Lipidomics_2009.pdf

下载此篇资料

J Chromatogr B Analyt Technol Biomed Life Sci. 2009 Sep 15;877(26):2664-72. doi: 10.1016/j.jchromb.2009.02.037. Epub 2009 Feb 24.

High-throughput shotgun lipidomics by quadrupole time-of-flight mass spectrometry.

Abstract

Technological advances in mass spectrometry and meticulous method development have produced several shotgun lipidomic approaches capable of characterizing lipid species by direct analysis of total lipid extracts. Shotgun lipidomics by hybrid quadrupole time-of-flight mass spectrometry allows the absolute quantification of hundreds of molecular glycerophospholipid species, glycerolipid species, sphingolipid species and sterol lipids. Future applications in clinical cohort studies demand detailed lipid molecule information and the application of high-throughput lipidomics platforms. In this review we describe a novel high-throughput shotgun lipidomic platform based on 96-well robot-assisted lipid extraction, automated sample infusion by mircofluidic-based nanoelectrospray ionization, and quantitative multiple precursor ion scanning analysis on a quadrupole time-of-flight mass spectrometer. Using this platform to compile comprehensive lipid arrays associated with metabolic dysfunctions is a powerful strategy for pinpointing the mechanistic details by which alterations in tissue-specific lipid metabolism are directly linked to the etiology of many lipid-mediated disorders.

 

2.2.  Automated sample infusion by mircofluidic-based nanoelectrospray ionization
Shotgun  lipidomics  has  been  successfully  performed  using nanoelectrospray   ionization   (nanoESI)   [19,21].   Conventional nanoESI  is  sensitive  and  cost-effective,  but  requires  manual loading  of  samples  into  small  glass  capillary  emitters,  manual positioning  of  emitters  in  front  of  the  mass  spectrometer  orifice  and  adjustment  of  voltage  and  back  pressure  to  achieve stable  ion  spray.  Automation  of  nanoESI  has  been  attempted by  continuous  flow  injection  for  analysis  of  multiple  samples [38]. Drawback of this technique is that all samples are infused through the same capillary emitter, which increases the risk for cross-contamination. An alternative is the automated chip-based nanoESI device: NanoMate TriVersa [39]. Here, the samples are aspirated robotically from a 96- or 384-well plate and infused into
the mass spectrometer through separate nozzles on an ESI Chip. Each ESI Chip contains 400 nozzles. ESI Chips are available with nozzles having an inner diameter of 2.5, 4.1 or 5.0 m. The device eliminates  cross-contamination  between  different  samples  and allows flow rates between 100 and 250 nL/min depending on ion spray  parameters;  e.g.  nozzle  diameter,  ion  spray  voltage,  back pressure and solvent composition.

 