Spectroscopic techniques have a great potential for analyzing
polymorphic forms and are used more and more frequently
for this kind of analysis. Especially IR and Raman spectroscopy offer fast and efficient identification and quantification capabilities. Mid IR spectroscopy, where fundamental
molecular vibrations are excited, is a routine method in the
Pharmacopoeias for the testing of the chemical identity. But
far IR spectroscopy, where back bone vibrations of larger
molecules or intermolecular vibrations in crystal lattices are
excited, is widely applied to the identification of polymorphic
forms and solvates.
The new Bruker FM functionality offers the measurement of
the complete FIR-MIR spectral range in one step and allows
for simultaneous identification and polymorph screening of
samples [1]. Nowadays, modern diamond ATR (attenuated
total reflection) accessories are used to analyze the substance directly without any sample preparation.
In the far IR region vacuum spectrometers are of high advantage. A vacuum far IR spectrometer eliminates disturbing
influences of atmospheric water vapor and increases the
stability of the measurement significantly while the sensitivity
is increased by a factor of two compared to non-vacuum far
IR [2].