特点
• Single-sample design
• Outstanding temperature stability
• Superior temperature uniformity verified by 40 matched thermocouples
• Optical encoders for the most accurate digital measurement of sample thickness
• International Standard Test Methods
ASTM C177 ISO 8302 DIN EN 12667
技术参数
技术
Single-Sample Advantages
Vertically Directed Heat Flow - Guaranteed
Superior Temperature Stability and NO Lateral Heat Flow
Uniform Temperature on both sample surfaces
仪器介绍:
The FOX GHP 600 is a Guarded Hot Plate system that provides absolute
thermal conductivity of insulating materials over a wide temperature
range in accordance with international standard test methods ASTM C177,
ISO 8302, and DIN EN 12667.
In a GHP system a temperature
difference (ΔT) is established across a sample of known thickness (L).
The thermal conductivity (l) is calculated from these values and the
steady state power (W) per area (A) required to maintain the temperature
difference. The Guarded Hot Plate uses a direct measurement of the
electrical power supplied to the hot plate. The LaserComp FOX GHP 600 is
especially well-suited for thermal conductivity measurements at high
temperatures, exhibiting unrivaled temperature and dimensional
stability, especially for measurements from 90 °C to 250 °C.